>> Rotorless Rheometer MFR 100Lite
|
Brief Descriptions:
|
Rotorless Rheometer MFR 100Lite is a newly-designed, in compliance with ASTM D5289, for testing the visco-elastic properties of rubber compounds. It's equipped with the latest temperature control system, advanced temperature probe and precise torque transducer as well as strict quality control in manufacture, which ensure MFR100Lite possessing excellent repeatability, reproducibility and sensitivity at the top level worldwide. Together with incorporating friendly graphical interfaces, powerful data processing capacity, and strong network function, it greatly improves users' testing capability, facility and efficiency.
|
Typical rotorless rheometer test-standard isothermal cure test:
|

|
S'—Elastic Torque; S"— Viscous Torque; TanDelta(Hysteresis Loss)—S"/S';
TC90 —90% Cure Time; TC50— 50% Cure Time;
ML— Minimum Torque; MH— Maximum Torque;
TS1— Scorch Time (time for 1 torque unit after ML); TS2— Scorch Time (time for 2 torque units after ML)
|
Application of typical isothermal cure test:
|
|
MFR TanDelta @ML corralates to Die Swell: MFR TanDelta @MH corralates to Rebond:

|
Repeatability and Reproducibility:
|
Test condition: EPDM (Lanxess Chemicals), Oscillation Angle: ±0.5°, 177℃, 100cpm (1.67Hz)
|
Data
|
S' @ML
|
S' @MH
|
TC10
|
TC50
|
TC90
|
S"@ML
|
S"@MH
|
Average value
|
1.40
|
11.59
|
1.41
|
3.07
|
5.31
|
0.91
|
1.02
|
Standard deviation
|
0.028
|
0.093
|
0.026
|
0.028
|
0.047
|
0.030
|
0.028
|
Repeatability difference coefficient
|
2.00
|
0.80
|
1.84
|
0.91
|
0.89
|
3.29
|
2.75
|
Standard deviation
|
0.069
|
0.13
|
0.078
|
0.089
|
0.173
|
0.088
|
0.086
|
Reproducibility difference coefficient
|
4.93
|
1.12
|
5.53
|
2.90
|
3.26
|
9.67
|
8.43
|
|
Technical Data and Specifications:
|
Standard
|
In compliance with ASTM D 5289 and GB-T 16584
|
Oscillating Angle
|
Standard: ±0.5°or ±0.2°or ±1.0°; Options: other Oscillation Angles customizable
|
Oscillation Frequency
|
100cpm (1.67Hz)
|
Sample Volume
|
About 4.5cm3
|
Temperature
|
Microprocessor Control
Temperature Range: room temperature ~200℃
Temperature Resolution: 0.01℃
Temperature Control Precision: ±0.1℃
Absolute Accuracy: ±0.2℃
Temperature Overheat: less than 0.4℃ at set point
Temperature Recovery Speed: less than 45 seconds
Temperature Rising Speed: less than 5 minutes (room temperature -100℃)
|
Data Measurement
|
Data type: Compound Torque S*, Elastic torque S', Viscous Torque S", Hysteresis Loss TanDelta,
any Scorch Time Point tsx, any Cure Time Point TCx,
Measuring Range: 0-200 dNm, measuring Precision: 0.01dNm
|
Test Type
|
Standard Cure Test (in compliance with ASTM D5289)
|
Calibration
|
Torque Spring Calibration
|
Display
|
7.5" TFT LCD, 640X480, 16 million pixel
|
Keyboard
|
64-key Membrane Keyboard
|
Embedded System
|
ARM9 Structure, 200MHz CPU, 64M Memory, 512M Onboard Storage, Linux Kernel
|
Interface
|
Fast Ethernet *1, USB 2.0*2, RS232*1
|
Electricity
|
220VAC±10%, 7A
|
Air Pressure
|
0.44MPa (4.5kg/cm2)
|
Size
|
Width: 68cm, Height: 128cm, Deep: 73cm
|
Weight
|
Gross Weight: 240kg; Net Weight: 190kg
|
Packing Size
|
L x W x H: 89cm x 80cm x 157cm
|
|